Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 21: CAL ADJ /5520 INSTRUMENT: Fluke 23: CAL ADJ /5520 INSTRUMENT: Fluke 73: CAL ADJ /5520 INSTRUMENT: Fluke 75: CAL ADJ /5520 INSTRUMENT: Fluke 77: CAL ADJ /5520 DATE: 19-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 60 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Fluke 70 Series service manual (PN 731034 Jan. 1984). # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R W 1.002 HEAD PRELIMINARY INSTRUCTIONS 1.003 DISP Warm-up time: NONE 1.003 DISP Temperature : 18 - 28 degC 1.003 DISP Humidity : <80% 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 HEAD CALIBRATION ACCESS PROCEDURE 1.006 DISP Select the V DC function (auto range). 1.006 DISP 1.006 DISP Remove the four phillips-head screws from the bottom of 1.006 DISP the case. Turn the instrument face-up, and grasp the top 1.006 DISP cover at both sides. Pull the top cover from the UUT. 1.007 HEAD CALIBRATION PROCEDURE 1.008 DISP Connect the 5520A and UUT as follows: 1.008 DISP [32] 5520A NORMAL HI to UUT Volt Ohm 1.008 DISP [32] 5520A NORMAL LO to UUT COM 1.009 5520 3V S 2W 1.010 DISP Adjust the Calibration Pot for a reading of 3.000V. 1.011 END